Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block.

Bigeleisen Paul E, Moayeri Nizar, Groen Gerbrand J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

A stimulation current of no more than 0.5 mA is regarded as safe in avoiding nerve injury and delivering adequate stimulus to provoke a motor response. However, there is no consistent level of stimulating threshold that reliably indicates intraneural placement of the needle. The authors determined the minimally required stimulation threshold to elicit a motor response outside and inside the most superficial part of the brachial plexus during high-resolution, ultrasound-guided, supraclavicular block.

METHODS

After institutional review board approval, ultrasound-guided, supraclavicular block was performed on 55 patients. Patients with neurologic dysfunction were excluded. Criteria for extraneural and intraneural stimulation were defined and assessed by independent experts. To determine success rate and any residual neurologic deficit, qualitative sensory and motor examinations were performed before and after block placement. At 6 month follow-up, the patients were examined for any neurologic deficit.

RESULTS

Thirty-nine patients met all set stimulation criteria. Median +/- SD (interquartile range) minimum stimulation threshold outside was 0.60 +/- 0.37 mA (0.40, 1.0) and inside 0.30 +/- 0.19 mA (0.20, 0.40). The difference of 0.30 mA was statistically significant (P < 0.0001). Stimulation currents of 0.2 mA or less were not observed outside the trunk in any patient. Significantly higher thresholds were observed in diabetic patients. Success rate was 100% after 20 min. Thirty-four patients had normal sensory and motor examination at 6 months. Five patients were lost to follow-up.

CONCLUSION

Within the limitations of this study and the use of ultrasound, a stimulation current of 0.2 mA or less is reliable to detect intraneural placement of the needle. Furthermore, stimulation currents of more than 0.2 and no more than 0.5 mA could not rule out intraneural position.

Для цитирования: Bigeleisen Paul E, Moayeri Nizar, Groen Gerbrand J. Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block.. Anesthesiology. 2009. DOI: 10.1097/aln.0b013e3181a59891.

Предметные рубрики MeSH