MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvs.2026.05.038

Медицинская библиотека · Journal of vascular surgery · 2026

Anatomic feasibility of off-the-shelf fenestrated endovascular aneurysm repair devices: An anatomic study.

Hofmann Amun, Leinweber Maria, Kliewer Miriam, Taher Fadi, Assadian Afshin, Plimon Markus

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Fenestrated endovascular aneurysm repair enables treatment of complex aortic aneurysms involving visceral branches but is limited by patient-specific anatomical variability. Off-the-shelf (OTS) devices may reduce manufacturing delay; however, their feasibility depends on anatomic compatibility across populations.

DESIGN

A large multicenter dataset was analyzed to evaluate the anatomic feasibility of standardized OTS fenestrated stent-graft configurations for juxtarenal and pararenal abdominal aortic aneurysms in this retrospective anatomic study.

METHODS

Computed tomography angiography data from a sample of 8238 patients evaluated for fenestrated endovascular aneurysm repair between January 2018 and December 2023 were analyzed. Measurements of the celiac axis (CA), superior mesenteric artery (SMA), left renal artery (LRA), and right renal artery (RRA) included ostial distance and branching angle relative to the CA. A greedy set-cover algorithm identified the smallest number of OTS configurations achieving fenestration target vessel alignment (FTVA). Cumulative coverage and clustering by aortic diameter were assessed. Analyses were exploratory and descriptive.

RESULTS

Anatomic data were available in 57.1% of cases for the SMA, 96.1% for the LRA, and 96.8% for the RRA. The CA served as the reference point for all analyses. Eight configurations covered 81.9% of SMA, 49.7% of LRA, and 52.9% of RRA anatomies. SMA coverage exceeded 90% with 16 designs, whereas 20 configurations were required for each renal artery to reach 80% coverage. No clustering of vessel origin by aortic diameter (median, 28 mm; interquartile range, 26-30.5 mm) was observed. Below 70% coverage for three-vessel and <25% for four-vessel combinations was achieved with the top 100 configurations.

CONCLUSIONS

A standardized dual-fenestrated OTS platform incorporating the CA and SMA appears anatomically feasible for most patients, whereas renal artery variability limits multivessel standardization. Dual-fenestrated devices with renal branches could also circumvent these constraints.

Для цитирования: Hofmann Amun, Leinweber Maria, Kliewer Miriam, Taher Fadi, Assadian Afshin, Plimon Markus. Anatomic feasibility of off-the-shelf fenestrated endovascular aneurysm repair devices: An anatomic study.. Journal of vascular surgery. 2026. DOI: 10.1016/j.jvs.2026.05.038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.