MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1142/s2424835520500460

Медицинская библиотека · The journal of hand surgery Asian-Pacific volume · 2020

A Novel Technique for Dynamic External Fixation of Proximal Interphalangeal Joint Fracture-Dislocations.

Sastravaha Nattawut, Limudomporn Kanon, Taweewuthisub Worasun

СтатьяПеревод в очереди

Аннотация

Background

The pins and rubbers traction system is widely used to treat unstable proximal interphalangeal (PIP) joint fracture-dislocations. However, dealing with rubber bands can be difficult due to its elasticity and weakness. The purpose of this study was to report the results of an alternative dynamic external fixator technique for unstable PIP joint injuries.

Methods

Thirteen patients with acute unstable PIP joint fracture-dislocation (n = 8), pilon fracture (n = 2), or comminuted shaft fractures of a middle phalanx with or without PIP joint extension (n = 3) were enrolled. The alternative fixator used a plastic syringe to maintain distraction force instead of the commonly used rubber bands. Four patients underwent combined open reduction due to unsatisfactory reduction by closed mean. The fixator was removed when there was radiographic evidence of healing.

Results

At 6 months after the operation, the mean active ROM of the injured PIP joint was 89° (range, 70-104°), and the mean Quick-DASH score was 9.78 (range, 0-40.9). The mean NRS of pain before fixator removal was 1 (range, 0-4). One patient had a pin tract infection and underwent debridement and fixator removal at 33 days postoperatively. There was no loss of reduction or implant failure in the study.

Conclusions

The results were comparable to other dynamic external fixation techniques used for unstable PIP joint fracture-dislocations. The main advantages of this technique were strength of the system, procedure simplicity, the compact design, and relatively low cost. In addition, this technique can also be used in pilon fractures and comminuted shaft fractures of a middle phalanx with or without PIP joint extension.

Для цитирования: Sastravaha Nattawut, Limudomporn Kanon, Taweewuthisub Worasun. A Novel Technique for Dynamic External Fixation of Proximal Interphalangeal Joint Fracture-Dislocations.. The journal of hand surgery Asian-Pacific volume. 2020. DOI: 10.1142/s2424835520500460.