Automated Detection and Removal of Pacing Artifacts from ECG Signal: CRIC and Cleveland Clinic Cohorts.

Dhar Rajkumar, Bukhari Hassaan A, Kewalramani Shivangi, Wong Claudia, Witzigreuter Luke, Rincon-Choles Hernan, Kansal Mayank, Hamm L Lee, He Jiang, Rao Panduranga, Bhat Zeenat, Gadegbeku Crystal, Jaar Bernard, Tereshchenko Larisa G, Anderson Amanda H, Appel Lawrence J, Chen Jing, Cohen Debbie L, Dember Laura M, Go Alan S, Lash James P, Rahman Mahboob

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Pacing artifacts in electrocardiogram (ECG) signals can interfere with waveform analysis and downstream quantitative interpretation. This study presents an algorithm for detecting and removing pacing artifacts from routine ECG recordings.

METHODS

The algorithm applies high pass filtering and Shannon energy computation to suppress cardiac components and enhance pacing artifacts. Principal component analysis (PCA) is then applied to the 12-lead ECGs to identify artifact start and end points from the first principal component. At each candidate location, slopes are calculated across leads, and artifacts are removed by linear interpolation only when the slope magnitude exceeds a predefined threshold. The algorithm was developed using 493 12-lead ECGs from the Chronic Renal Insufficiency Cohort and validated using 203 Cleveland Clinic 12-lead ECGs, 80 Cardiac Memory with ICD 12 lead ECGs, and 200 10-second two-lead Holter ECG epochs from the MIT-BIH Arrhythmia Database. A graphical user interface was developed for review and refinement.

RESULTS

The algorithm automatically detected and removed pacing artifacts in 84% of validation ECG files; the remaining files required user-guided refinement. Overall sensitivity and specificity for pacing artifact detection were 98.4% and 76.2%, respectively.

CONCLUSION

The proposed algorithm detects and removes pacing artifacts with substantial automation while allowing guided refinement in challenging recordings.

SIGNIFICANCE: The framework combines multilead PCA-based localization, lead-specific slope confirmation, and adaptive artifact-width removal directly on routine ECG, enabling morphology-preserving preprocessing for quantitative and AI-enabled ECG analysis.

Для цитирования: Dhar Rajkumar, Bukhari Hassaan A, Kewalramani Shivangi, Wong Claudia, Witzigreuter Luke, Rincon-Choles Hernan, Kansal Mayank, Hamm L Lee, He Jiang, Rao Panduranga, Bhat Zeenat, Gadegbeku Crystal, Jaar Bernard, Tereshchenko Larisa G, Anderson Amanda H, Appel Lawrence J, Chen Jing, Cohen Debbie L, Dember Laura M, Go Alan S, Lash James P, Rahman Mahboob. Automated Detection and Removal of Pacing Artifacts from ECG Signal: CRIC and Cleveland Clinic Cohorts.. IEEE transactions on bio-medical engineering. 2026. DOI: 10.1109/tbme.2026.3716891.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.