Feasibility of Continuous Intraoperative Neural Monitoring During Transoral Endoscopic Thyroidectomy Vestibular Approach in a Porcine Model.

Zhang Daqi, Li Shijie, Dionigi Gianlorenzo, Wang Tie, Zhang Jiao, Xue Gaofeng, Sun Hui

СтатьяПеревод в очереди

Аннотация

Background

Experimental study to evaluate feasibility of continuous intraoperative neural monitoring (C-IONM) in transoral endoscopic thyroidectomy vestibular approach (TOETVA).

Methods

Duroc-Landrace pigs were orally intubated with electromyogram endotracheal tube. Automatic periodic stimulation (APS) electrode was allocated in the operative field through the 5-mm ports. APS was then repose on vagal nerve (VN) with different approaches: (1) median, that is, between sternothyroid and thyroid gland; and (2) lateral, that is, between sternocleidomastoid and sternothyroid. VN was stimulated proximally and distally to the APS location to verify whether the dissection and/or placement determined VN injury. Video presentation is offered.

Results

Assembled APS accessory was feasible in large-brained animals. The two basic options for VN approaches were tested. Baseline obtained had amplitude values >1000 μV, bilaterally.

Conclusions

C-IONM was feasible in TOETVA in porcine models, but simplification of electrode design and application is needed.

Для цитирования: Zhang Daqi, Li Shijie, Dionigi Gianlorenzo, Wang Tie, Zhang Jiao, Xue Gaofeng, Sun Hui. Feasibility of Continuous Intraoperative Neural Monitoring During Transoral Endoscopic Thyroidectomy Vestibular Approach in a Porcine Model.. Journal of laparoendoscopic & advanced surgical techniques. Part A. 2018. DOI: 10.1089/lap.2018.0054.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.