In vitro biomechanical evaluation and comparison of a new prototype locking plate and a limited-contact self compression plate for equine fracture repair.

Quinteros D D, García-López J M, Abuja G A, Tarkanian M, Maranda L S, Bubeck K, Kowaleski M P

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To determine if the mechanical properties (strength and stiffness) of a new prototype 4.5 mm broad locking plate (NP-LP) are comparable with those of a traditional 4.5 mm broad limited-contact self compression plate (LC-SCP), and to compare the bending and torsional properties of the NP-LP and LC-SCP when used in osteotomized equine third metacarpal bones (MC3).

METHODS

The plates alone were tested in four-point bending single cycle to failure. The MC3-plate constructs were created with mid-diaphyseal osteotomies with a 1 cm gap. Constructs were tested in four-point bending single cycle to failure, four-point bending cyclic fatigue, and torsion single cycle to failure.

RESULTS

There were not any significant differences in bending strength and stiffness found between the two implants. The MC3-NP-LP construct was significantly stiffer than the MC3-LC-SCP in bending. No other biomechanical differences were found in bending, yield load in torsion, or mean composite rigidity. Mean cycles to failure for bending fatigue testing were similar for both constructs.

CLINICAL SIGNIFICANCE: The NP-LP was comparable to the LC-SCP in intrinsic, as well as structural properties. The NP-LP construct was more rigid than the LC-SCP construct under four-point bending, and both constructs behaved similarly under four-point bending cyclic fatigue testing and torsion single cycle to failure. The new NP-LP implant fixation is biomechanically comparable to the LC-SCP in a simulated MC3 fracture.

Для цитирования: Quinteros D D, García-López J M, Abuja G A, Tarkanian M, Maranda L S, Bubeck K, Kowaleski M P. In vitro biomechanical evaluation and comparison of a new prototype locking plate and a limited-contact self compression plate for equine fracture repair.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2012. DOI: 10.3415/vcot-11-01-0011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.