OBJECTIVE

Axillary Intra-aortic balloon pump (IABP) has been increasingly utilized for hemodynamic support in heart failure patients. Vascular complications associated with axillary IABP such as dissection or rupture are relatively rare but not negligible that could negatively affect clinical outcomes. We summarized our experiences.

METHODS

This is a retrospective study reviewing of all patients receiving an axillary IABP between June 2016 and November 2020. A total of 199 patients underwent percutaneous axillary IABP placement. 6 patients (6/199, 3.0%) were complicated with arterial/aortic dissection or rupture during the procedures or the course of treatment. We described their clinical presentations and outcomes.

RESULTS

Vascular complications included acute type A aortic dissection in 2 patients, descending aortic rupture in 1 patient, abdominal aortic rupture along with type B aortic dissection in 1 patient, and the localized left subclavian artery dissection in 2 patient. 2 type A aortic dissection cases were surgically treated: 1 with emergent left ventricle assist device and ascending aorta replacement, the other with emergent left ventricle assist device. Emergent endovascular treatment was successfully performed in 2 aortic rupture cases. The left subclavian artery dissection cases were managed medically. The postoperative/treatment course was uneventful in all patients.

CONCLUSION

Percutaneous axillary IABP therapy can cause significant vascular complications. Early diagnosis and prompt treatment would be the key to improve the clinical outcomesv.

Для цитирования: Nishida Hidefumi, Song Tae, Onsager David, Nguyen Ann, Grinstein Jonathan, Chung Bow, Smith Bryan, Kalantari Sara, Sarswat Nitasha, Kim Gene, Pinney Sean, Jeevanandam Valluvan, Milner Ross, Ota Takeyoshi. Significant vascular complications in percutaneous axillary intra-aortic balloon pump.. Annals of vascular surgery. 2022. DOI: 10.1016/j.avsg.2021.12.078.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.