Evaluation of Thermal, Hematohistological, and Dermatological Biocompatibility of LED Devices for Neonatal Phototherapy.

Nascimento Tayomara Ferreira, Bocchi Silvia Cristina Mangini, Lyra João Cesar, Bianchi Rodrigo Fernando, Junior Lauro de Assis Duarte, Lacerda Giselle Silveira, Abadde Luciana Patrícia Fernandes, Rocha Noeme Sousa, Vieira Susana Eduardo, Langoni Hélio, Nascimento Cristiano Neves do, Jensen Rodrigo

СтатьяПеревод в очереди

Аннотация

Background/

Objective

The effectiveness of blue-light phototherapy (PT) is mainly dependent on the total dose of light (time under PT and amount of skin exposed) received by infants. The primary aim of this study was the development of a novel, flexible, and stretchable device to provide continuous PT treatment, avoiding temporary interruptions that are often observed in practice, such as during breastfeeding, for example. This study evaluated the biocompatibility of a novel, low-cost blanket equipped with light-emitting diode (LED) lamps designed to maintain therapeutic efficacy while facilitating uninterrupted skin-to-skin contact.

Methods

Fourteen New Zealand White rabbits, weighing approximately 2.9 kg and aged 4 months, were randomly assigned to an experimental group (TG, n = 7) or a control group (CG, n = 7). The TG received phototherapy directly on the skin (irradiance: 19.3 [13.0-22.0] µW/cm-2/nm-1) during two 12 h sessions over consecutive days, while the CG remained under identical conditions with the device turned off. Biochemical, hematological, dermatological, and histological parameters, as well as rectal and skin temperatures, were assessed.

Results

The results showed no differences in clinical appearance or histological analysis of skin tissue between the groups. Blood analysis indicated a reduction in absolute monocyte counts in the TG compared to the CG (p = 0.049), though levels remained within normal ranges. Skin temperature was consistently higher in the TG, except during the initial measurement. Rectal temperatures were similar on the first day but lower in the TG on the second day (mean 40.3 ± 0.21 °C vs. 40.7 ± 0.32 °C; p = 0.039).

Conclusions

Temperature levels remained within physiological limits for both groups throughout the study. The device demonstrated biocompatibility and caused no adverse dermatological, hematological, or biochemical effects.

Для цитирования: Nascimento Tayomara Ferreira, Bocchi Silvia Cristina Mangini, Lyra João Cesar, Bianchi Rodrigo Fernando, Junior Lauro de Assis Duarte, Lacerda Giselle Silveira, Abadde Luciana Patrícia Fernandes, Rocha Noeme Sousa, Vieira Susana Eduardo, Langoni Hélio, Nascimento Cristiano Neves do, Jensen Rodrigo. Evaluation of Thermal, Hematohistological, and Dermatological Biocompatibility of LED Devices for Neonatal Phototherapy.. Biomedicines. 2025. DOI: 10.3390/biomedicines13112826.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.