DOI 10.1016/j.jfma.2023.11.003
Polypropylene anterior-apical single-incision UPHOLD-LITE mesh surgery in women with severe pelvic organ prolapse: Outcome at 53 months follow up.
Аннотация
BACKGROUND AND PURPOSE: We present this current study to complement with mesh inlays plausible benefits (UPHOLD-LITE System) on available long-term study amidst FDA's concern on mesh complications. This study aims to assess the medium-term outcomes of UPHOLD-LITE system for treatment of advanced pelvic organ prolapse (POP) and its complications, and lower urinary tract symptoms.
METHODS
This is a retrospective case series of 53 months follow-up of 123 consecutive patients who underwent UPHOLD-LITE system. Objective outcome measures the anatomical correction of prolapse with POP-Q ≤ Stage 1. Subjective outcome was patient's feedback to questions 2 and 3 of POPDI-6. Secondary outcome measures the quality of life, presence of lower urinary tract symptoms and complications. Quality of life is assessed by validated questionnaires on Urogenital Distress Inventory 6 (UDI-6), Incontinence Impact Questionnaire 7 (IIQ-7), Pelvic Organ Prolapse Distress Inventory 6 (POPDI-6), and Pelvic Organ Prolapse/Urinary Incontinence Sexual Function Questionnaire 12 (PISQ-12) at 1 and 3 years post-operatively.
RESULTS
Objective outcome at 1 and 3 years was at 96.7 % and 95.4 % respectively. The subjective cure was 95.1 % and 91.6 %. Five-year cumulative cure rate maintained at 87.2 %. Secondary outcomes observed improvement on UDI-6, IIQ-7, POPDI-6 and PISQ-12 postoperatively. Bladder outlet obstruction improved while de novo urodynamic stress incontinence (USI) increased slightly post surgically. Mesh erosion rate was 0.8 %.
CONCLUSION
The UPHOLD-LITE system demonstrated good medium term anatomical correction of apical and anterior prolapse, with good subjective cure and improved quality of life. Whilst complication rate was low, slight increase in de novo USI was observed.