DOI 10.1002/hed.26334

Медицинская библиотека · Head & neck · 2020

Safety parameters of ferromagnetic device during thyroid surgery: Porcine model using continuous neuromonitoring.

Huang Tzu-Yen, Lin Yi-Chu, Tseng Hsin-Yi, Dionigi Gianlorenzo, Kim Hoon Yub, Chai Young Jun, Lu I-Cheng, Chang Pi-Ying, Chiang Feng-Yu, Wu Che-Wei

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The ferromagnetic (FM) device is a new surgical energy modality. This study investigated dynamic recurrent laryngeal nerve (RLN) electromyographic (EMG) data to define safety parameters for using FM devices in thyroidectomy.

METHODS

Real-time EMG tracings obtained under continuous neuromonitoring were recorded from 24 RLNs (12 piglets). In the activation study, FM devices were activated at varying distances from the RLN. In the cooling study, FM devices were cooled for varying time intervals, or muscle touch maneuver was performed, before contact with the RLN.

RESULTS

During the FMwand/FMsealer activation, no adverse EMG events occurred at distances of 2 mm or longer. The cooling study revealed no adverse EMG events after 1-second (FMwand) or 3-second (FMsealer) intervals or after muscle touch maneuver.

CONCLUSIONS

An FM device should be activated at a distance of 2 mm from the RLN and should be adequately cooled before further contact with the RLN. Surgeons can avoid RLN injury by observing standard procedures for using FM devices.

Для цитирования: Huang Tzu-Yen, Lin Yi-Chu, Tseng Hsin-Yi, Dionigi Gianlorenzo, Kim Hoon Yub, Chai Young Jun, Lu I-Cheng, Chang Pi-Ying, Chiang Feng-Yu, Wu Che-Wei. Safety parameters of ferromagnetic device during thyroid surgery: Porcine model using continuous neuromonitoring.. Head & neck. 2020. DOI: 10.1002/hed.26334.