Performance and Safety of the Extravascular Implantable Cardioverter Defibrillator Through Long-Term Follow-Up: Final Results From the Pivotal Study.
СтатьяМногоцентровое исследованиеКлиническое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Friedman Paul, Murgatroyd Francis, Boersma Lucas V A, Manlucu Jaimie, Knight Bradley P, Clémenty Nicolas, Leclercq Christophe, Amin Anish, Merkely Béla, Birgersdotter-Green Ulrika Maria, Chan Joseph Yat Sun, Biffi Mauro, Knops Reinoud Elwin, Engel Gregory, Muñoz Carvajal Ignacio, Epstein Laurence M, Sagi Venkata, Johansen Jens Brock, Sterliński Maciej, Steinwender Clemens, Hounshell Troy, Abben Richard, Thompson Amy E, Zhang Yan, Wiggenhorn Christopher, Willey Sarah, Crozier Ian, Extravascular ICD Pivotal Study Investigators
Аннотация
BACKGROUND: Substernal lead placement of the extravascular implantable cardioverter defibrillator (EV ICD) permits both defibrillation at thresholds similar to those seen with transvenous implantable cardioverter defibrillators and effective anti-tachycardia pacing (ATP) while avoiding the vasculature and associated complications. The global Pivotal study has shown the EV ICD system to be safe and effective through 6 months, but long-term experience has yet to be published. Our aim was to report the performance and safety of the EV ICD system throughout the study.
METHODS: The EV ICD Pivotal study was a prospective, global, single-arm, premarket clinical study. Individuals with a Class I or IIa indication for a single-chamber implantable cardioverter defibrillator per guidelines were enrolled. Freedom from major system- or procedure-related complications and appropriate and inappropriate therapy rates were assessed through 3 years with the Kaplan-Meier method. ATP success was calculated from simple proportions.
RESULTS: An implantation was attempted in 316 patients (25.3% female; 53.8±13.1 years of age; 81.6% primary prevention; left ventricular ejection fraction, 38.9±15.4%). Of 299 patients with a successful implantation, 24 experienced 82 spontaneous arrhythmic episodes that were appropriately treated with ATP only (38, 46.3%), shock only (34, 41.5%), or both (10, 12.2%) for a Kaplan-Meier-estimated rate of first any appropriate therapy of 9.2% at 3 years. ATP was successful in 77.1% (37/48) of episodes, and ATP use significantly increased from discharge to last follow-up visit (P<0.0001). Shock therapy was successful in 100% (27/27) of discrete, spontaneous ventricular arrhythmias. The inappropriate shock rates at 1 and 3 years were 9.8% and 17.5%, respectively, with P-wave oversensing the predominant cause. No major intraprocedural complications were reported, and the estimated freedom from system- or procedure-related major complications was 91.9% at 1 year and 89.0% at 3 years. The most common major complications were lead dislodgement (10 events; n=9 patients, 2.8%), postoperative wound or device pocket infection (n=8, 2.5%), and device inappropriate shock delivery (n=4, 1.3%). Twenty-four system revisions were performed as a result of major complications related to the EV ICD system or procedure.
CONCLUSIONS: From implantation to study completion, the EV ICD Pivotal study demonstrated that a single integrated system with an extravascular lead placed in the substernal space maintains high ATP success, effective defibrillation, and a consistent safety profile.
REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT04060680.
METHODS: The EV ICD Pivotal study was a prospective, global, single-arm, premarket clinical study. Individuals with a Class I or IIa indication for a single-chamber implantable cardioverter defibrillator per guidelines were enrolled. Freedom from major system- or procedure-related complications and appropriate and inappropriate therapy rates were assessed through 3 years with the Kaplan-Meier method. ATP success was calculated from simple proportions.
RESULTS: An implantation was attempted in 316 patients (25.3% female; 53.8±13.1 years of age; 81.6% primary prevention; left ventricular ejection fraction, 38.9±15.4%). Of 299 patients with a successful implantation, 24 experienced 82 spontaneous arrhythmic episodes that were appropriately treated with ATP only (38, 46.3%), shock only (34, 41.5%), or both (10, 12.2%) for a Kaplan-Meier-estimated rate of first any appropriate therapy of 9.2% at 3 years. ATP was successful in 77.1% (37/48) of episodes, and ATP use significantly increased from discharge to last follow-up visit (P<0.0001). Shock therapy was successful in 100% (27/27) of discrete, spontaneous ventricular arrhythmias. The inappropriate shock rates at 1 and 3 years were 9.8% and 17.5%, respectively, with P-wave oversensing the predominant cause. No major intraprocedural complications were reported, and the estimated freedom from system- or procedure-related major complications was 91.9% at 1 year and 89.0% at 3 years. The most common major complications were lead dislodgement (10 events; n=9 patients, 2.8%), postoperative wound or device pocket infection (n=8, 2.5%), and device inappropriate shock delivery (n=4, 1.3%). Twenty-four system revisions were performed as a result of major complications related to the EV ICD system or procedure.
CONCLUSIONS: From implantation to study completion, the EV ICD Pivotal study demonstrated that a single integrated system with an extravascular lead placed in the substernal space maintains high ATP success, effective defibrillation, and a consistent safety profile.
REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT04060680.