m+Medelevator.RU
← Поиск публикаций
Archives of disease in childhood. Fetal and neonatal edition · 2023

Apnoea-triggered increase in fraction of inspired oxygen in preterm infants: a randomised cross-over study.

Рандомизированное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Marshall Andrew, Ladlow Oliver J, Bannink Charlotte, Lim Kathleen, Ali Sanoj K M, Gale Timothy J, Dargaville Peter A

Аннотация

OBJECTIVES: To investigate the impact of a pre-emptive apnoea triggered oxygen response on oxygen saturation (SpO2) targeting following central apnoea in preterm infants.

DESIGN: Interventional crossover study of a 12-hour period of automated oxygen control with an apnoea response (AR) module, nested within a crossover study of a 24-hour period of automated oxygen control compared with aggregated data from two flanking 12-hour periods of manual control.

SETTING: Neonatal intensive care unit PATIENTS: Preterm infants receiving non-invasive respiratory support and supplemental oxygen; median (IQR) birth gestation 27 (26-28) weeks, postnatal age 17 (12-23) days.

INTERVENTION: Automated oxygen titration with an automated control algorithm modified to include an AR module. Alterations to inspired oxygen concentration (FiO2) were actuated by a motorised blender. Desired SpO2 range was 90-94%. Apnoea detection was by capsule pneumography.

MAIN OUTCOME MEASURES: Duration, magnitude and area under the curve (AUC) of SpO2 deviations following apnoea; frequency and duration of apnoeic events. Comparisons between periods of manual, automated and automated control with AR module.

RESULTS: In 60 studies in 35 infants, inclusion of the AR module significantly reduced AUC for SpO2 deviations below baseline compared with both automated and manual control (manual: 87.1%±107.6% s, automated: 84.6%±102.8% s, AR module: 79.4%±102.7% s). However, there was a coincident increase in SpO2 overshoot (AUC (SpO2>SpO2(onset)); manual: 44.3±99.9% s, automated: 54.7%±103.4% s, AR module: 65.7%±126.2% s).

CONCLUSION: Automated control with a pre-emptive apnoea-triggered FiO2 boost resulted in a modest reduction in post-apnoea hypoxaemia, but was followed by a greater SpO2 overshoot.

TRIAL REGISTRATION NUMBER: ACTRN12616000300471.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты