Apnoea-triggered increase in fraction of inspired oxygen in preterm infants: a randomised cross-over study.

Marshall Andrew, Ladlow Oliver J, Bannink Charlotte, Lim Kathleen, Ali Sanoj K M, Gale Timothy J, Dargaville Peter A

Рандомизированное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

To investigate the impact of a pre-emptive apnoea triggered oxygen response on oxygen saturation (SpO2) targeting following central apnoea in preterm infants.

DESIGN

Interventional crossover study of a 12-hour period of automated oxygen control with an apnoea response (AR) module, nested within a crossover study of a 24-hour period of automated oxygen control compared with aggregated data from two flanking 12-hour periods of manual control.

SETTING

Neonatal intensive care unit PATIENTS: Preterm infants receiving non-invasive respiratory support and supplemental oxygen; median (IQR) birth gestation 27 (26-28) weeks, postnatal age 17 (12-23) days.

INTERVENTION: Automated oxygen titration with an automated control algorithm modified to include an AR module. Alterations to inspired oxygen concentration (FiO2) were actuated by a motorised blender. Desired SpO2 range was 90-94%. Apnoea detection was by capsule pneumography.

MAIN OUTCOME MEASURES: Duration, magnitude and area under the curve (AUC) of SpO2 deviations following apnoea; frequency and duration of apnoeic events. Comparisons between periods of manual, automated and automated control with AR module.

RESULTS

In 60 studies in 35 infants, inclusion of the AR module significantly reduced AUC for SpO2 deviations below baseline compared with both automated and manual control (manual: 87.1%±107.6% s, automated: 84.6%±102.8% s, AR module: 79.4%±102.7% s). However, there was a coincident increase in SpO2 overshoot (AUC (SpO2>SpO2(onset)); manual: 44.3±99.9% s, automated: 54.7%±103.4% s, AR module: 65.7%±126.2% s).

CONCLUSION

Automated control with a pre-emptive apnoea-triggered FiO2 boost resulted in a modest reduction in post-apnoea hypoxaemia, but was followed by a greater SpO2 overshoot.

TRIAL REGISTRATION NUMBER: ACTRN12616000300471.

Для цитирования: Marshall Andrew, Ladlow Oliver J, Bannink Charlotte, Lim Kathleen, Ali Sanoj K M, Gale Timothy J, Dargaville Peter A. Apnoea-triggered increase in fraction of inspired oxygen in preterm infants: a randomised cross-over study.. Archives of disease in childhood. Fetal and neonatal edition. 2023. DOI: 10.1136/archdischild-2023-325849.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.