Evaluating the Efficacy of Cervical Tactile Ultrasound Technique as a Predictive Tool for Spontaneous Preterm Birth.
Аннотация
BACKGROUND
Premature cervical softening and shortening may be considered an early mechanical failure that predispose to preterm birth.
PURPOSE: This study aims to explore the applicability of an innovative cervical tactile ultrasound approach for predicting spontaneous preterm birth (sPTB).
MATERIALS AND
METHODS
Eligible participants were women with low-risk singleton pregnancies in their second trimester, enrolled in this prospective observational study. A Cervix Monitor (CM) device was designed with a vaginal probe comprising four tactile sensors and a single ultrasound transducer operating at 5 MHz. The probe enabled the application of controllable pressure to the external cervical surface, facilitating the acquisition of stress-strain data from both anterior and posterior cervical sectors. Gestational age at delivery was recorded and compared against cervical elasticity.
RESULTS
CM examination data were analyzed for 127 women at 240/7 - 286/7 gestational weeks. sPTB was observed in 6.3% of the cases. The preterm group exhibited a lower average cervical stress-to-strain ratio (elasticity) of 0.70 ± 0.26 kPa/mm compared to the term group's 1.63 ± 0.65 kPa/mm with a p-value of 1.1 × 10-4. Diagnostic accuracy for predicting spontaneous preterm birth based solely on cervical elasticity data was found to be 95.0% (95% CI, 88.5 - 100.0).
CONCLUSION
These findings suggest that measuring cervical elasticity with the designed tactile ultrasound probe has the potential to predict spontaneous preterm birth in a cost-effective manner.