MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fped.2025.1569437

Медицинская библиотека · Frontiers in pediatrics · 2025

Evaluation of a novel reprocessed bCPAP system on sepsis rates among preterm neonates with respiratory distress: a randomized controlled trial.

Badin Sarah, Roodaki Navid, Garcia Daisy Evangeline C, Abila-Cariaga Rochelle, Burke Thomas F

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Bubble CPAP (bCPAP) is highly effective in the treatment of respiratory distress syndrome of prematurity and other causes of newborn respiratory insufficiency. To overcome barriers to bCPAP access a novel system was developed that is designed to be cleaned, disinfected, and reused. This study evaluated whether use of reprocessed bCPAP systems increases the rate of sepsis in neonates.

METHODS

A post hoc analysis of a single-center randomized controlled trial (registration no. NCT06082674) was conducted that compared mechanical ventilator driven CPAP devices (MV-CPAP) with single-use circuits to reusable bCPAP systems that were cleaned and disinfected after each use. The primary outcome was a composite of treatment escalation or death.

RESULTS

Seventy-five neonates were randomized to the two CPAP treatment arms. No significant differences in death (5 vs. 4), escalation of care (10 vs. 9), and the composite outcome (OR = 0.84; 95% CI: 0.30-2.35, p = 0.743) were detected in the MV-CPAP and bCPAP groups respectively. There were no clinically significant differences in any of the secondary outcomes.

DISCUSSION: Use of a reprocessed bCPAP system designed to increase global access to CPAP did not increase rates of neonatal sepsis.

CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov, identifier, (NCT06082674).

Для цитирования: Badin Sarah, Roodaki Navid, Garcia Daisy Evangeline C, Abila-Cariaga Rochelle, Burke Thomas F. Evaluation of a novel reprocessed bCPAP system on sepsis rates among preterm neonates with respiratory distress: a randomized controlled trial.. Frontiers in pediatrics. 2025. DOI: 10.3389/fped.2025.1569437.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.