PURPOSE: Fetoscopic laser coagulation for twin-to-twin transfusion syndrome is challenging for anterior placenta due to the rigidity of current tools. The capacity to keep entry port forces minimal is critical for this procedure, as is optimal coagulation distance and orientation. This work introduces technological tools to this end.

METHODS

A novel fetoscope is presented with a rigid shaft and a flexible steerable segment at the distal end. The steerable segment can bend up to 90[Formula: see text] even when loaded with a laser fiber. An artificial pneumatic muscle makes such acute bending possible while allowing for a low-weight and disposable device.

RESULTS

The flexible fetoscope was validated in a custom-made phantom model to measure visual range and coagulation efficacy. The flexible fetoscope shows promising results when compared to a clinical rigid curved fetoscope to reach anterior targets. The new fetoscope was then evaluated in vivo (pregnant ewe) where it successfully coagulated placental vasculature.

CONCLUSION

The flexible fetoscope improved the ability to achieve optimal coagulation angle and distance on anteriorly located targets. The fetoscope also showed the potential to lead fetoscopic laser coagulation and other fetal surgical procedures toward safer and more effective interventions.

Для цитирования: Ahmad Mirza Awais, Ourak Mouloud, Wenmakers Dirk, Valenzuela Ignacio, Basurto David, Ourselin Sebastien, Vercauteren Tom, Deprest Jan, Poorten Emmanuel Vander. Development and validation of a flexible fetoscope for fetoscopic laser coagulation.. International journal of computer assisted radiology and surgery. 2023. DOI: 10.1007/s11548-023-02905-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.