MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/1749-799x-4-10

Медицинская библиотека · Journal of orthopaedic surgery and research · 2009

In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prostheses.

Pastrav Leonard C, Jaecques Siegfried Vn, Jonkers Ilse, Perre Georges Van der, Mulier Michiel

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The per-operative assessment of primary stem stability may help to improve the performance of total hip replacement. Vibration analysis methods have been successfully used to assess dental implant stability, to monitor fracture healing and to measure bone mechanical properties. The objective of the present study was to evaluate in vivo a vibration analysis-based endpoint criterion for the insertion of the stem by successive surgeon-controlled hammer blows.

METHODS

A protocol using a vibration analysis technique for the characterisation of the primary bone-prosthesis stability was tested in 83 patients receiving a custom-made, intra-operatively manufactured stem prosthesis. Two groups were studied: one (n = 30) with non cemented and one (n = 53) with partially cemented stem fixation. Frequency response functions of the stem-femur system corresponding to successive insertion stages were compared.

RESULTS

The correlation coefficient between the last two frequency response function curves was above 0.99 in 86.7% of the non cemented cases. Lower values of the final correlation coefficient and deviations in the frequency response pattern were associated with instability or impending bone fracture. In the cases with a partially cemented stem an important difference in frequency response function between the final stage of non cemented trial insertion and the final cemented stage was found in 84.9% of the cases. Furthermore, the frequency response function varied with the degree of cement curing.

CONCLUSION

The frequency response function change provides reliable information regarding the stability evolution of the stem-femur system during the insertion. The protocol described in this paper can be used to accurately detect the insertion end point and to reduce the risk for intra-operative fracture.

Для цитирования: Pastrav Leonard C, Jaecques Siegfried Vn, Jonkers Ilse, Perre Georges Van der, Mulier Michiel. In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prostheses.. Journal of orthopaedic surgery and research. 2009. DOI: 10.1186/1749-799x-4-10.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.