[Comparison of Shikani optical stylet and Macintosh laryngoscope for double-lumen endotracheal tube intubation].
Аннотация
OBJECTIVE
To compare the efficacy and safety of Shikani (S)optical stylet and Macintosh(M) laryngoscope for double-lumen endotracheal tube intubation.
METHODS
In the study, 60 patients undergoing elective thoracic surgery were randomly allocated to group S(n=30) and group M (n=30). After general anesthesia induction, the patients in group S and group M were intubated double-lumen endotracheal tube (DLT) by Shikani optical stylet (SOS) and macintosh laryngoscope respectively. Intubation time, intubation attempts, cuff broken and oral mucosal or dental injury were recorded; Blood pressure and heart rate at baseline (T0), at the time of intubaiton onset (T1), 1 minute after intubaiton (T2), 3 minutes after intubation (T3) and 5 minutes after intubation (T3) were also recorded; Hoarseness and throat sore of the patients 24 hours after surgery were evaluated.
RESULTS
The intubaiton time with the SOS was faster than with the Macintosh [(37.4±9.7) s vs. (43.9±13.7) s, P=0.039] and the first attempt success rate (87% vs. 80%, P=0.488) did not differ between the groups; No tube cuff broke in both the groups; Group S had fewer patients who suffered oral mucosal or dental injury than group M (8 vs.2, P=0.038); The blood pressure and heart rate at T0,T1,T2,T3 and T4 did not differ between the groups; Throat sore (7 vs.10, P=0.390) and hoarseness (5 vs.7, P=0.519) incidence did not differ between the groups.
CONCLUSION
By comparison of the Macintosh laryngoscope, the SOS provides faster DLT intubation and causes less oral Mucosal or dental injury.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.