DOI 10.1016/0029-7844(96)00227-x

Медицинская библиотека · Obstetrics and gynecology · 1996

Skull fracture caused by vacuum extraction.

Hickey K, McKenna P

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

The vacuum extractor is being increasingly advocated as the instrument of first choice for assisted vaginal delivery. It is widely believed that the vacuum cup will dislodge before causing serious fetal trauma.

CASE: Rotational delivery of a term infant was effected using a vacuum extractor. A 6-cm Malmström metal cup with a paramedian application was in place for 12 minutes. The vacuum pressure developed was 0.8 kg/cm2. Four traction efforts with contractions were required to deliver the fetal head. A neonatal skull x-ray the following day showed a comminuted parietal bone fracture at the vacuum cup application site. Management was conservative, and the infant's neurologic behavior remained normal.

CONCLUSION

The vacuum extractor exerts considerable traction force. Fetal skull fracture can result, and its true incidence may be higher than expected, considering that few neonates with normal neurologic behavior undergo skull x-ray.

Для цитирования: Hickey K, McKenna P. Skull fracture caused by vacuum extraction.. Obstetrics and gynecology. 1996. DOI: 10.1016/0029-7844(96)00227-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.