DOI 10.1016/j.healun.2024.05.004
Long-term outcomes of a novel fully magnetically levitated ventricular assist device for the treatment of advanced heart failure in China.
Аннотация
BACKGROUND
Left ventricular assist devices (LVADs) are well-established for treating end-stage heart failure, but this therapy is only available to Chinese patients in recent years. The CH-VAD is the first used fully magnetically levitated pump in China. This study reports the long-term outcomes of a cohort supported by the CH-VAD for the first time.
METHODS
From June 2017 to August 2023, 50 consecutive patients received CH-VAD implantation in Fuwai Hospital. Clinical data were collected and retrospectively analyzed.
RESULTS
Baseline characteristics included a mean age of 47.9 ± 13.9 years, 90% male, and 26% ischemic etiology. The Interagency Registry for Mechanically Assisted Circulatory Support profile revealed 12% profile 1, 56% profile 2, 26% profile 3, and 6% profile 4. The mean support duration was 868 ± 630 days (range 33 days-6.4 years). Kaplan-Meier survival rate was 93% (95% CI, 79-98) at 1 year, 93% (95% CI, 79-98) at 2 years, and 89% (95% CI, 71-96) at 3 years. Forty patients (80%) currently remain on support, 3 were bridged to recovery, 2 received transplants, and 5 expired during support. Major adverse events (AEs) included right heart failure (10%), surgical-related bleeding (8%), arrhythmia (8%), and driveline infection (16%). Major hemocompatibility-related AEs were limited to 3 nondisabling strokes and 1 gastrointestinal bleeding. No major device malfunction occurred during the follow-up period.
CONCLUSIONS
The largest single-center experience with the leading LVAD in China shows high survival with low complication rates, demonstrating that CH-VAD is safe and efficient in providing long-term support for patients with end-stage heart failure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.