Point-of-care measurements of blood ketones in newborns.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Harris Deborah L, Weston Philip J, Harding Jane E
Аннотация
OBJECTIVE: Babies may use alternative cerebral fuels including ketones when blood glucose concentrations are low, but laboratory ketone measurements are slow and expensive. Point-of-care measurement of ketone concentrations, if sufficiently accurate, may provide useful information for clinical care.
PATIENTS AND DESIGN: Eligible babies were 35-42 weeks' gestation, ≤10 days old and admitted to the newborn intensive care unit. At the time of clinically indicated blood tests, additional samples were taken to measure beta-hydroxybutyrate using a point-of-care analyser and the laboratory method.
RESULTS: One-hundred and fifty babies had 142 paired samples. Overall point-of-care accuracy was excellent (mean difference 0.00 mmol/L) and precision was moderate (SD 0.18 mmol/L). A point-of-care measurement ≥0.4 mmol/L was highly predictive of a laboratory measurement ≥0.4 mmol/L (area under the curve 0.98).
CONCLUSION: Point-of-care measurement of blood beta-hydroxybutyrate concentrations is sufficiently accurate in newborns to be potentially useful in clinical care.
CLINICAL TRIAL REGISTRATION NUMBER: Registered with the Australian and New Zealand Clinical Trials Registry ACTRN: 12616000784415. The study was registered before recruitment commenced.
PATIENTS AND DESIGN: Eligible babies were 35-42 weeks' gestation, ≤10 days old and admitted to the newborn intensive care unit. At the time of clinically indicated blood tests, additional samples were taken to measure beta-hydroxybutyrate using a point-of-care analyser and the laboratory method.
RESULTS: One-hundred and fifty babies had 142 paired samples. Overall point-of-care accuracy was excellent (mean difference 0.00 mmol/L) and precision was moderate (SD 0.18 mmol/L). A point-of-care measurement ≥0.4 mmol/L was highly predictive of a laboratory measurement ≥0.4 mmol/L (area under the curve 0.98).
CONCLUSION: Point-of-care measurement of blood beta-hydroxybutyrate concentrations is sufficiently accurate in newborns to be potentially useful in clinical care.
CLINICAL TRIAL REGISTRATION NUMBER: Registered with the Australian and New Zealand Clinical Trials Registry ACTRN: 12616000784415. The study was registered before recruitment commenced.