Point-of-care measurements of blood ketones in newborns.
Аннотация
OBJECTIVE
Babies may use alternative cerebral fuels including ketones when blood glucose concentrations are low, but laboratory ketone measurements are slow and expensive. Point-of-care measurement of ketone concentrations, if sufficiently accurate, may provide useful information for clinical care.
PATIENTS AND
DESIGN
Eligible babies were 35-42 weeks' gestation, ≤10 days old and admitted to the newborn intensive care unit. At the time of clinically indicated blood tests, additional samples were taken to measure beta-hydroxybutyrate using a point-of-care analyser and the laboratory method.
RESULTS
One-hundred and fifty babies had 142 paired samples. Overall point-of-care accuracy was excellent (mean difference 0.00 mmol/L) and precision was moderate (SD 0.18 mmol/L). A point-of-care measurement ≥0.4 mmol/L was highly predictive of a laboratory measurement ≥0.4 mmol/L (area under the curve 0.98).
CONCLUSION
Point-of-care measurement of blood beta-hydroxybutyrate concentrations is sufficiently accurate in newborns to be potentially useful in clinical care.
CLINICAL TRIAL REGISTRATION NUMBER: Registered with the Australian and New Zealand Clinical Trials Registry ACTRN: 12616000784415. The study was registered before recruitment commenced.