Development of in-vitro pulsatile flow generator for evaluating the performance of hemodialysis catheters.
Аннотация
Hemodialysis (HD) using an HD catheter is performed widely on renal failure patients. The catheter was evaluated using the recirculation ratio in pre-clinical status, which is a crucial index indicating its performance. However, pre-clinical in-vivo experiments have limitations: high cost, and ethical issues. Hence, computational and in-vitro methods have been developed as alternatives. However, computational methods require fluid dynamic knowledge, whereas in-vitro experiments are complicated and expensive. In this study, we developed a pulsatile flow generator to mimic blood flow achieving cost effectiveness and user convenience. The device used iterative learning control, achieving blood flow in the superior and inferior vena cava within a 3.3% error. Furthermore, the recirculation ratios were measured based on two insertion directions and two different external pipe materials to evaluate the catheter regarding patients' posture and blood vessel stiffness. The results provide a better understanding of cardiovascular device performance without complicated and costly pre-clinical tests.