Is skin pressure a relevant factor for socket assessment in patients with lower limb amputation?
Аннотация
BACKGROUND
Prosthetic rehabilitation improves the overall quality of life of patients, despite discomfort and medical complications. No quantitative assessment of prosthesis-patient interaction is used in routine protocols and prosthesis quality still results from the manufacturer's know-how.
OBJECTIVE
Our objective is to investigate whether pressure can be a relevant factor for assessing socket adequacy.
METHODS
A total of 8 transtibial amputee volunteers took part in this experimental study. The protocol included static standing and 2 minutes walking tests while the stump-to-socket interface pressures were measured. Questionnaires on comfort and pain were also conducted.
RESULTS
During static standing test, maximum pressures were recorded in the proximal region of the leg, with a peak value reaching 121.1 ± 31.6 kPa. During dynamic tests, maximum pressures of 254.1 ± 61.2 kPa were recorded during the loading phase of the step. A significant correlation was found between the pain score and static maximum recorded pressure (r= 0.81).
CONCLUSIONS
The protocol proposed and evaluated in this study is a repeatable, easy-to-set quantified analysis of the patient to socket interaction while standing and walking. This approach is likely to improve feedback for prosthesis manufacturers and consequently the overall design of prostheses.