DOI 10.1046/j.1537-2995.1996.36111297091744.x

Медицинская библиотека · Transfusion · 1996

Evaluation of swirling, pH, and glucose tests for the detection of bacterial contamination in platelet concentrates.

Wagner S J, Robinette D

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Although infrequent, episodes of transfusion-associated bacterial sepsis may lead to serious outcomes or death and therefore are of concern. This study evaluates the sensitivity of three surrogate tests for the presence of bacteria in platelet concentrates: cessation of swirling, low extracellular pH, and low plasma glucose levels.

STUDY DESIGN AND

METHODS

Day 0 platelet concentrates were inoculated with low levels of one of seven bacterial strains and stored with agitation at 20 to 24 degrees C for up to 5 days. In the morning and afternoon of each day of storage, bacterial levels were ascertained by quantitative plate culture, and platelet concentrates were tested for platelet pH, plasma glucose, and swirling. Quantitative and semiquantitative dipstick techniques were used to determine pH and glucose levels.

RESULTS

Platelet concentrates had attained stationary phase growth of Streptococcus pyogenes, Staphylococcus aureus, Escherichia coli, Bacillus cereus, Enterobacter cloacae, Streptococcus mitis, or Staphylococcus epidermidis (> or = 10(7) -10(8) colony-forming units [CFU]/mL) before either the swirling or dipstick methods suggested the presence of bacteria. The sensitivity of swirling and glucose tests for detecting bacterial contamination in platelet concentrates was generally comparable. Although the sensitivity of the pH test was generally similar to that of swirling and glucose tests for most bacteria, platelet concentrates contaminated with E. cloacae had normal pH despite the presence of high levels of bacteria (> or = 10(7)-10(8) CFU/mL).

CONCLUSION

The sensitivity in detection of bacteria in platelet concentrates by the swirling technique or by measuring extracellular pH or plasma glucose is less than the sensitivity reported for microscopy with Gram stain (10(5)-10(6) CFU/mL), fluorescence microscopy with acridine orange stain (10(4)-10(5) CFU/mL), chemiluminescence detection of ribosomal RNA (10(3)-10(4) CFU/mL), and automated bacterial culture (1 CFU/sample volume).

Для цитирования: Wagner S J, Robinette D. Evaluation of swirling, pH, and glucose tests for the detection of bacterial contamination in platelet concentrates.. Transfusion. 1996. DOI: 10.1046/j.1537-2995.1996.36111297091744.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.