Randomised In Vitro Study Investigating PEEP-Stability During Application of CPAP With Binasal Prongs and Face Masks.
Аннотация
AIM: Face masks and binasal prongs are commonly used interfaces for applying continuous positive airway pressure (CPAP) in neonatology. We aimed to assess CPAP stability in a randomised controlled in vitro study.
METHODS
In a simulated resuscitation scenario of a 1000-g preterm infant with respiratory distress, 20 operators (10 with/without neonatology experience) aimed to maintain a CPAP of 5 cmH2O as precisely as possible using face masks or binasal prongs in random order. The primary outcome was the minimum-achieved CPAP at the Y-piece (PYmin). Secondary outcomes included time to target CPAP, CPAP stability and the impact of operator experience.
RESULTS
Binasal prongs enabled more consistent maintenance of the target CPAP of 5 cmH2O than face masks (median [IQR]: PYmin: binasal prongs, 3.74 cmH2O [3.54-3.88]) vs. 3.20 cmH2O (2.72-3.73), with no significant deviation from the target CPAP. Target CPAP was achieved significantly faster with face masks (2.89 s [1.67-5.64] vs. 6.49 s [4.76-13.62] [p < 0.05]). No significant differences were observed on the basis of operator experience (p > 0.05).
CONCLUSION
Binasal prongs allow more accurate CPAP maintenance than face masks regardless of the operator's experience, although current clinical studies offer limited evidence on the superiority of nasal interfaces compared with that of face masks.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.