Аннотация

BACKGROUND

The sliding hip screw (SHS) is frequently used in the management of hip fractures; successful placement depends on accurate positioning of the lag screw in the femoral head guided by fluoroscopy. We proposed to leverage the capabilities of augmented reality (AR) to overlay virtual images of the desired guidewire trajectory directly onto the surgical field to guide the surgeon during SHS guidewire insertion.

METHODS

Using a commercially available AR headset and software, we performed preprocedural planning using computed tomography scans to identify the optimal trajectory for SHS guidewire insertion in the neck of a Sawbones femur model. The images of the scanned femurs containing the virtual guidewire trajectory were overlaid on the physical models such that the user could see a composite view of the computer-generated images and the physical environment. Two second-year orthopedic residents each inserted 15 guidewires under AR guidance and 15 guidewires under fluoroscopy.

RESULTS

Of the 30 guidewires inserted under AR guidance, 24 (80%) were within the femoral neck, and 16 (53%) were fully enclosed within the femoral head. Nine (56%) of the 16 perforations were due to insertions that were too far along the planned trajectory. Thirteen (81%) of the successful attempts with AR had an appropriate position, compared to 25/26 (96%) with fluoroscopy. It took significantly less time to perform the procedure using fluoroscopy than AR (p < 0.05). Fluoroscopy required on average 18.7 shots.

CONCLUSION

Augmented reality provides an opportunity to aid in guidewire insertion in a preplanned trajectory with less radiation exposure in a sterile environment, but technical challenges remain to be solved to enable widespread adoption.

Для цитирования: Laverdière Carl, Corban Jason, Ge Susan, Kang Yukyung, Harvey Edward, Martineau Paul A, Noel Geoffroy, Reindl Rudolf. Augmented-reality-guided insertion of sliding hip screw guidewire: a preclinical investigation.. Canadian journal of surgery. Journal canadien de chirurgie. 2022. DOI: 10.1503/cjs.025620.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.