PURPOSE: Pericardial effusion (PE) may occur during the lead extraction procedure (TLE). Little is known about the incidence, causes, and predictors of this complication.

METHODS

From January 2009 to October 2016, TLE was attempted for 297 leads in 212 patients (age 69.3 ± 12.9 years, 169 male, BMI 27.2 ± 9.9 m²/kg, LVEF 43.4 ± 24.6%) for lead dysfunction (62.7%), upgrade (16.0%), infection (14.2%), or other (7.0%) indications. TLE was performed under general anesthesia with continuous invasive arterial blood pressure and transesophageal echocardiography (TEE) monitoring. For lead removal, the mechanical approach was first attempted, followed by the laser-assisted technique when needed. Severity of PE was defined by the presence of hemodynamically significant PE > 10 mm at TEE.

RESULTS

Clinical success was achieved for 292 leads (98.3%). New-onset PE was observed in 14 patients (6.6%) [mild entity in 7 patients (3.3%) and severe in 7 (3.3%)]. In these latter patients, intra-procedural management included surgery (n = 3), pericardiocentesis (n = 2), or a conservative approach (n = 2). Right ventricular (RV) site lesions were treated with a simple fluid infusion. Laceration of the superior vena cava and other vessels resulted in rescue surgery. Lesions of the right atrial free wall (n = 1) and coronary sinus (n = 1) were treated with pericardiocentesis. NYHA III/IV, LVEF < 35%, renal impairment, right-sided implant, and ≥2 leads targeted for TLE were associated with new-onset PE. More than two factors identified a higher risk group (16.2%, 95% CI 6.2-32.0%, P = 0.02).

CONCLUSIONS

New-onset PE is common during TLE and is associated with specific factors. PE severity and subsequent patient management depend on the site of injury.

Для цитирования: Regoli François, D'Ambrosio Gabriele, Caputo Maria Luce, Svab Stefano, Conte Giulio, Moccetti Tiziano, Klersy Catherine, Cassina Tiziano, Demertzis Stefanos, Auricchio Angelo. New-onset pericardial effusion during transvenous lead extraction: incidence, causative mechanisms, and associated factors.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2018. DOI: 10.1007/s10840-018-0327-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.