Three-dimensional rotational X-ray imaging for spine surgery: a quantitative validation study comparing reconstructed images with corresponding anatomical sections.

Verlaan Jorrit-Jan, van de Kraats Everine B, van Walsum Theo, Dhert Wouter J A, Oner F Cumhur, Niessen Wiro J

Сравнительное исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

STUDY

DESIGN

A validation study was done in which reconstructed three-dimensional rotational x-ray images were quantitatively compared with corresponding anatomic sections.

OBJECTIVES

To assess the accuracy of reconstructed images acquired on a three-dimensional rotational x-ray imaging device.

SUMMARY OF BACKGROUND DATA: Minimally invasive procedures have proven quite successful as alternatives for a multitude of open treatments. An unfavorable property of this type of procedure is a lack of direct vision of the operating area. Three-dimensional rotational x-ray imaging may be able to merge the advantages of computed tomography and fluoroscopy: real-time two-dimensional projections for fast visual feedback and three-dimensional reconstructions for detailed volumetric imaging of complex anatomy.

METHODS

Twenty traumatic thoracolumbar burst fractures were created and underwent pedicle screw instrumentation and balloon vertebroplasty. Subsequently, a three-dimensional dataset was obtained, and the midsagittal image was reconstructed. The specimens were sliced, and photographs were obtained. Multiple parameters on the reconstructed images and photographs were measured two times by two observers using a graphical method. The differences and standard deviations were calculated for the corresponding parameters and for the intraobserver data.

RESULTS

The mean difference between the corresponding values ranged between -1.1 and 2.1 mm for all parameters. The standard deviation for the differences per parameter ranged between 1.2 and 3.2 mm. The intraobserver differences ranged from -0.8 to 1.4 mm, and the standard deviation varied between 0.4 and 2.4 mm.

CONCLUSIONS

The reconstructed midsagittal images were accurate in all measured parameters. The three-dimensional rotational x-ray technique may prove to be valuable for less invasive spine surgery.

Для цитирования: Verlaan Jorrit-Jan, van de Kraats Everine B, van Walsum Theo, Dhert Wouter J A, Oner F Cumhur, Niessen Wiro J. Three-dimensional rotational X-ray imaging for spine surgery: a quantitative validation study comparing reconstructed images with corresponding anatomical sections.. Spine. 2005. DOI: 10.1097/01.brs.0000154650.31781.85.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.