Аннотация

PURPOSE: To develop a device for percutaneous transrenal ureteral occlusion.

MATERIALS AND

METHODS

The device was a double-body Gianturco-Rösch biliary stent constrained at the junction of the two stents to create an hourglass shape. One stent was coated with silicone. One device was percutaneously placed in each of nine pigs through a 9-F Teflon sheath. Urographic and hematologic follow-up was performed for up to 12 weeks.

RESULTS

Seven pigs showed immediate, complete ureteral occlusion, and two pigs exhibited persistent incomplete high-grade obstruction. All animals exhibited varying degrees of hydronephrosis and hydroureter. No device migration was noted. Minor complications were encountered during device placement in three pigs. Mucosal folds and villus-like projections that arose from the lamina propria protruded into the lumen of the ureter at the cranial end of the covered stent and around the wire of the caudal stent. Varying degrees of mural inflammation and edema were noted.

CONCLUSION

Transrenal ureteral occlusion with the described device appears to be a viable method for treating urinary fistulas.

Для цитирования: Marr B, Wright K C, Carrasco C H, Brewer L. Experimental evaluation of a new device for percutaneous transrenal ureteral occlusion.. Journal of vascular and interventional radiology : JVIR. 1997. DOI: 10.1016/s1051-0443(97)70659-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.