High-throughput monitoring of in vitro enzymatic reactions using probe electrospray ionization-mass spectrometry.

Okahashi Nobuyuki, Taniguchi Takeo, Suzuki Takuma, Hattori Takanari, Takahashi Hidenori, Honda Kohsuke, Matsuda Fumio

СтатьяПеревод в очереди

Аннотация

Probe electrospray ionization-mass spectrometry (PESI-MS) is an ambient ionization technique that enables rapid and simple mass spectrometry with minimal sample preparation. In this study, we evaluated the applicability of PESI-MS for the quantitative analysis of metabolites in in vitro enzyme reaction mixtures, which often contain high concentrations of proteins, salts, and buffers that are not suitable for conventional MS. Using glycolytic metabolites as a model, we optimized the type of organic solvents for mixing, data processing method, and investigated the effects of additives and reaction components on ionization efficiency and signal stability. Ethanol and 2-propanol were identified as suitable solvents for the detection of sugar phosphates, organic acids, and nucleotides. The quantitative reproducibility was improved by employing a total peak-height-based data processing method. The presence of proteins had a minimal influence on ionization. Although high concentrations of MgCl2 and buffers suppressed the signal intensity, metabolite signals were still observable at 10 and 50 mM, respectively. The linearity of the calibration curves and the reproducibility of the analysis of glycolytic metabolites in the mock enzyme reaction solution were confirmed. Finally, PESI-MS was successfully used to monitor the time course of intermediates in an in vitro synthetic non-oxidative glycolysis system. The data trend obtained by PESI-MS was roughly consistent with the results of conventional LC-MS analysis. These findings demonstrate that PESI-MS is a simple, rapid, and robust method for analyzing in vitro enzymatic reactions.

Для цитирования: Okahashi Nobuyuki, Taniguchi Takeo, Suzuki Takuma, Hattori Takanari, Takahashi Hidenori, Honda Kohsuke, Matsuda Fumio. High-throughput monitoring of in vitro enzymatic reactions using probe electrospray ionization-mass spectrometry.. Analytical and bioanalytical chemistry. 2026. DOI: 10.1007/s00216-026-06656-2.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.