DOI 10.1177/1558944719877890

Медицинская библиотека · Hand (New York, N.Y.) · 2019

A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size.

Ilyas Asif M, Mahoney Jonathan M, Bucklen Brandon S

СтатьяПеревод в очереди

Аннотация

Background

There is evidence that interfragmentary fracture gap size may affect the compression achievable with a modern headless compression screw (HCS). This mechanical study compared the compression achieved by 3 commercial HCS systems through various fracture gaps: CAPTIVATE Headless (Globus Medical, Inc, Audubon, Pennsylvania), Synthes (DePuy Synthes, Westchester, Pennsylvania), and Acumed Acutrak 2 (Acumed LLC, Hillsboro, Oregon).

Methods

Screws were inserted into a custom test fixture composed of polyurethane synthetic bone foam fragments, separated by a layer of easily compressible polyurethane foam simulating a fracture gap. Compression was measured after final insertion and countersinking. The effect of the interfragmentary fracture gap size on the compression generated was also investigated.

Results

The CAPTIVATE Headless 3.0 mm screw (70.1 ± 5.7 N) and the Synthes 3.0 mm screw (64.9 ± 7.3 N) achieved similar compressive forces after final countersink. Similar comparisons were found for the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws, and the CAPTIVATE Headless 4.0 mm and Acutrak 2 Standard screws. The final compression of the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws was not significantly affected when the fracture gap was doubled from 2 to 4 mm, but was reduced significantly by 95.9% with the Acutrak 2 Micro screw.

Conclusion

When comparing like-sized screws, the CAPTIVATE, Synthes, and Acutrak 2 HCS systems demonstrated similar potential compressive forces. However, compared with the CAPTIVATE Headless and Synthes HCS systems, which are inserted with a compression sleeve that is not gap distance-dependent, the Acutrak 2 HCS system demonstrated less compression when the simulated fracture gap size was increased to 4 mm.

Для цитирования: Ilyas Asif M, Mahoney Jonathan M, Bucklen Brandon S. A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size.. Hand (New York, N.Y.). 2019. DOI: 10.1177/1558944719877890.