Spinal accessory neuropathy following dual tunneling of lead extensions for motor cortex stimulation.
Аннотация
BACKGROUND
Spinal accessory nerve (SAN) injury is a known complication of surgery involving the posterior cervical triangle. Because the nerve lies in a plane that is traversed by extension leads for neuromodulation, the SAN is at risk during tunneling for these procedures.
METHODS
We report a 50-year-old man who developed spinal accessory neuropathy after tunneling of two sets of extension leads through his neck to connect to the pulse generator at the chest wall. Immediately postoperatively, he developed severe right shoulder pain and trapezius weakness manifested as decreased shoulder abduction and scapular winging.
RESULTS
Electromyography 2 months later confirmed a SAN injury. Ultrasonography demonstrated the SAN to be angulated between the two sets of extension leads in the posterior cervical triangle. Operative exploration confirmed this finding. Neurolysis was performed and the deep extension lead was relocated superficial to the nerve. Postoperatively, the patient had immediate resolution of his shoulder pain and an improved range of motion.
CONCLUSION
As the indications for neuromodulation continue to grow, tunneling may be performed more frequently. It is our belief that the second pass through the neck of a dual extension lead system may inadvertently be directed deeply and poses an increased risk to the SAN.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.