Atlantoaxial fusion: a biomechanical analysis of two C1-C2 fusion techniques.

Claybrooks Roderick, Kayanja Mark, Milks Ryan, Benzel Edward

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Different atlantoaxial fusion techniques are used for instability. Transarticular screws are biomechanically superior to wiring techniques and equivalent to C1 lateral mass to C2 pedicle (C1LM-C2P) fixation. Recently, C1 lateral mass to C2 laminar (C1LM-C2L) fixation has been shown to have flexibility similar to C1LM-C2P fixation in flexion, extension, lateral bending, and axial rotation.

PURPOSE: Compare the stiffness of C1LM-C2P with C1LM-C2L screw rod fixation.

STUDY

DESIGN

In vitro biomechanical study.

OUTCOME MEASURES: Stiffness in flexion/extension, lateral bending, axial rotation, and anterior-posterior (AP) translation.

METHODS

Eight fresh-frozen human cadaveric cervical spines (C1-C3) were tested intact and, after a type II odontoid fracture, were instrumented and tested with two fixation constructs: C1LM-C2P screws and C1LM-C2L screws. The testing involved flexion, extension, lateral bending, AP translation, and axial rotation. Stiffness was measured and compared with a repeated-measures analysis.

RESULTS

C1LM-C2P was significantly stiffer than the intact in AP translation (p<.001), lateral bending (p=.001), and axial rotation (p=.002) and equivalent in flexion/extension (p=.09). C1LM-C2L was significantly stiffer than the intact in AP translation (p<.01) and axial rotation (p<.004) and equivalent in lateral bending (p<.71) and flexion/extension (p=.22). C1LM-C2P was stiffer than C1LM-C2L in right/left lateral bending (p<.001) and axial rotation (p=.009) and equivalent in AP translation (p=.06) and flexion/extension (p=.74).

CONCLUSION

C1LM-C2P fixation is equivalent to C1LM-C2L fixation in flexion/extension and AP translation and superior in lateral bending and axial rotation.

Для цитирования: Claybrooks Roderick, Kayanja Mark, Milks Ryan, Benzel Edward. Atlantoaxial fusion: a biomechanical analysis of two C1-C2 fusion techniques.. The spine journal : official journal of the North American Spine Society. 2007. DOI: 10.1016/j.spinee.2006.08.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.