Arthroscopic transosseous rotator cuff repair: how to avoid damaging the axillary nerve-a cadaveric study.
Аннотация
INTRODUCTION
Arthroscopic transosseous rotator cuff repair can be performed with an external guide, although the proximity to the axillary nerve raises safety concerns. The aim of this study is to determine the safety of different drilling angles regarding the axillary nerve.
MATERIALS AND
METHODS
We performed a bone tunnel in the greater tuberosity in 17 fresh frozen shoulders, using an external guide at four different angles: 40°, 50°, 60°, and 70°. At each angle, we measured the distance between the drill and the axillary nerve, the distance from the acromion to the skin incision point, and the perimeter of the arm at the axilla.
RESULTS
The distance to the axillary nerve was safe with the guide at an angle of 40°, 50° and 60°, but not at 70° (p = 0.001). We found significant differences between all four angles (p < 0.05). Regression analysis demonstrated the influence of the guide angle in all measurements assessed (p < 0.001). There was no association between the measurements taken and the axillary perimeter (p > 0.5).
CONCLUSIONS
Arthroscopic transosseous rotator cuff repair with an external guide does not pose a risk for the axillary nerve using angles of 60° or less.