Аннотация

OBJECTIVE

In this study we report on finite element calculation to verify the mechanical stability induced by the new geometry of a novel condylar process plate. We report our clinical experiences in the treatment of condylar process fractures using the novel plate. The plate has been designed according to finite element calculations in order to optimize mechanical stability and stiffness. The potential influence of this plate on the postoperative complication rate was assessed.

STUDY

DESIGN

Thirty-five mandibular condylar process fractures in 30 patients (20 male, 10 female, age: 34.3 +/- 11.5 years) were treated with the Modus condylar process plates (Medartis, Basle, Switzerland) and are included in this study. The time periods for postoperative follow-ups were 6 weeks and 6 months.

RESULTS

No plate fractures occurred and there was no injury to the facial nerve. In one patient a screw loosening was observed. Two patients showed inadequate reduction without functional impairment.

CONCLUSION

The newly developed condylar process plate proved to provide sufficient mechanical stiffness to avoid plate fractures. The postoperative complication rate can be reduced significantly. Statistical comparison to other osteosynthesis methods of the condylar process was performed and showed significant differences (single miniplates: P = .006 and minidynamic compression plates: P = .055). We conclude that a single Modus condylar process plate shows results comparable to 2 miniplates.

Для цитирования: Seemann Rudolf, Schicho Kurt, Reichwein Astrid, Eisenmenger G, Ewers Rolf, Wagner Arne. Clinical evaluation of mechanically optimized plates for the treatment of condylar process fractures.. Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics. 2007. DOI: 10.1016/j.tripleo.2007.07.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.