Adedinsewo Demilade, Morales-Lara Andrea Carolina, Hardway Heather, Johnson Patrick, Young Kathleen A, Garzon-Siatoya Wendy Tatiana, Butler Tobah Yvonne S, Rose Carl H, Burnette David, Seccombe Kendra, Fussell Mia, Phillips Sabrina, Lopez-Jimenez Francisco, Attia Zachi I, Friedman Paul A, Carter Rickey E, Noseworthy Peter A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Cardiomyopathy is a leading cause of pregnancy-related mortality and the number one cause of death in the late postpartum period. Delay in diagnosis is associated with severe adverse outcomes.

OBJECTIVE

To evaluate the performance of an artificial intelligence-enhanced electrocardiogram (AI-ECG) and AI-enabled digital stethoscope to detect left ventricular systolic dysfunction in an obstetric population.

METHODS

We conducted a single-arm prospective study of pregnant and postpartum women enrolled at 3 sites between October 28, 2021, and October 27, 2022. Study participants completed a standard 12-lead ECG, digital stethoscope ECG and phonocardiogram recordings, and a transthoracic echocardiogram within 24 hours. Diagnostic performance was evaluated using the area under the curve (AUC).

RESULTS

One hundred women were included in the final analysis. The median age was 31 years (Q1: 27, Q3: 34). Thirty-eight percent identified as non-Hispanic White, 32% as non-Hispanic Black, and 21% as Hispanic. Five percent and 6% had left ventricular ejection fraction (LVEF) <45% and <50%, respectively. The AI-ECG model had near-perfect classification performance (AUC: 1.0, 100% sensitivity; 99%-100% specificity) for detection of cardiomyopathy at both LVEF categories. The AI-enabled digital stethoscope had an AUC of 0.98 (95% CI: 0.95, 1.00) and 0.97 (95% CI: 0.93, 1.00), for detection of LVEF <45% and <50%, respectively, with 100% sensitivity and 90% specificity.

CONCLUSION

We demonstrate an AI-ECG and AI-enabled digital stethoscope were effective for detecting cardiac dysfunction in an obstetric population. Larger studies, including an evaluation of the impact of screening on clinical outcomes, are essential next steps.

Для цитирования: Adedinsewo Demilade, Morales-Lara Andrea Carolina, Hardway Heather, Johnson Patrick, Young Kathleen A, Garzon-Siatoya Wendy Tatiana, Butler Tobah Yvonne S, Rose Carl H, Burnette David, Seccombe Kendra, Fussell Mia, Phillips Sabrina, Lopez-Jimenez Francisco, Attia Zachi I, Friedman Paul A, Carter Rickey E, Noseworthy Peter A. Artificial intelligence-based screening for cardiomyopathy in an obstetric population: A pilot study.. Cardiovascular digital health journal. 2024. DOI: 10.1016/j.cvdhj.2024.03.005.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.