MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1071100718806474

Медицинская библиотека · Foot & ankle international · 2018

Risk Factors for Failure of Total Ankle Arthroplasty With a Minimum Five Years of Follow-up.

Cody Elizabeth A, Bejarano-Pineda Lorena, Lachman James R, Taylor Michel A, Gausden Elizabeth B, DeOrio James K, Easley Mark E, Nunley James A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

: As the popularity of total ankle arthroplasty (TAA) increases and indications expand, surgeons require a better understanding of which patient factors are associated with implant failure. In this study, we aimed to use a large total ankle database to identify independent risk factors for implant failure at mid- to long-term follow-up.

METHODS

: A prospectively collected database was used to identify all patients who underwent primary TAA with a minimum 5 years' follow-up. The primary outcome was revision, defined as removal of one or both metal components; failures due to infection were excluded. Patient and clinical factors analyzed included age, sex, body mass index (BMI), smoking status, presence of diabetes, indication for TAA, implant, tourniquet time, and presence of ipsilateral hindfoot fusion. Preoperative coronal deformity and sagittal talar translation were assessed, as were postoperative coronal and sagittal tibial component alignment. Univariable and multivariable analyses were performed to identify predictors of implant failure. After excluding 5 ankles that failed because of deep infection, 533 ankles with a mean 7 (range, 5-11) years of follow-up met the inclusion criteria. Four implants were used: INBONE I, INBONE II, STAR, and Salto-Talaris.

RESULTS

: Thirty-four ankles (6.4%) were revised or removed a mean 4 (range, 1-9) years postoperatively. The only independent predictors of failure were the INBONE I prosthesis and ipsilateral hindfoot fusion ( P = .006 and P = .023, respectively).

CONCLUSIONS

: This is among the largest studies to analyze the relationship between TAA failure rates and multiple different patient, operative, and radiographic factors. Of note, age, BMI, and amount of deformity were not associated with higher failure rates. Only patients with ipsilateral hindfoot fusion or who received the INBONE I prosthesis were at significantly higher risk of implant failure.

LEVEL OF EVIDENCE:: Level III, retrospective cohort study.

Для цитирования: Cody Elizabeth A, Bejarano-Pineda Lorena, Lachman James R, Taylor Michel A, Gausden Elizabeth B, DeOrio James K, Easley Mark E, Nunley James A. Risk Factors for Failure of Total Ankle Arthroplasty With a Minimum Five Years of Follow-up.. Foot & ankle international. 2018. DOI: 10.1177/1071100718806474.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.