MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/aor.70222

Медицинская библиотека · Artificial organs · 2026

Experimental Extracorporeal Circulation System: A Pilot Study in Rats.

Kawaguchi Akira T, Inukai Kaori, Shibata Mayuko, Kawaguchi Gen T, Sato Kazuaki, Aburaya Taiga, Bergsland Jacob, Yamada Yoshiyuki

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Extracorporeal circulation (ECC) has become an important component in cardiac surgery, for cardiopulmonary support, and as a bridge to transplantation. Optimization of systems, components, and management requires further refinements using relevant ECC models.

METHODS

A miniature silicone membrane oxygenator with a surface area of 134 cm2 and a pump-head tube with internal dimension, ID 2.0 mm, was connected to the rat with a double-lumen tube, where the inner silicone tube (ID 1.0 mm) was irrigated with temperature-controlled water, serving as a heat-exchanger.

RESULTS

The ECC system had a priming volume less than 1.5 mL yet developed a pulsatile flow up to 11.6 mL/min, or 40 mL/min/kg, for a 290-g animal. The system kept sufficient gas exchange, sustained stable hemodynamics in heart failure, and induced hyperthermia (40°C) and hypothermia (20°C) successfully with long-term survival.

CONCLUSION

A silicone-based closed ECC system for rodents was developed as an experimental platform to address controversies, optimize management, and test the feasibility of new systems, components, or concepts in ECC.

Для цитирования: Kawaguchi Akira T, Inukai Kaori, Shibata Mayuko, Kawaguchi Gen T, Sato Kazuaki, Aburaya Taiga, Bergsland Jacob, Yamada Yoshiyuki. Experimental Extracorporeal Circulation System: A Pilot Study in Rats.. Artificial organs. 2026. DOI: 10.1111/aor.70222.