Semiautomated optical coherence tomography-guided robotic surgery for porcine lens removal.
СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Chen Cheng-Wei, Francone Anibal Andrés, Gerber Matthew J, Lee Yu-Hsiu, Govetto Andrea, Tsao Tsu-Chin, Hubschman Jean-Pierre
Аннотация
PURPOSE: To evaluate semiautomated surgical lens extraction procedures using the optical coherence tomography (OCT)-integrated Intraocular Robotic Interventional Surgical System.
SETTING: Stein Eye Institute and Department of Mechanical and Aerospace Engineering, University of California, Los Angeles, USA.
DESIGN: Experimental study.
METHODS: Semiautomated lens extraction was performed on postmortem pig eyes using a robotic platform integrated with an OCT imaging system. Lens extraction was performed using a series of automated steps including robot-to-eye alignment, irrigation/aspiration (I/A) handpiece insertion, anatomic modeling, surgical path planning, and I/A handpiece navigation. Intraoperative surgical supervision and human intervention were enabled by real-time OCT image feedback to the surgeon via a graphical user interface. Manual preparation of the pig-eye models, including the corneal incision and capsulorhexis, was performed by a trained cataract surgeon before the semiautomated lens extraction procedures. A scoring system was used to assess surgical complications in a postoperative evaluation.
RESULTS: Complete lens extraction was achieved in 25 of 30 eyes. In the remaining 5 eyes, small lens pieces (≤1.0 mm3) were detected near the lens equator, where transpupillary OCT could not image. No posterior capsule rupture or corneal leakage occurred. The mean surgical duration was 277 seconds ± 42 (SD). Based on a 3-point scale (0 = no damage), damage to the iris was 0.33 ± 0.20, damage to the cornea was 1.47 ± 0.20 (due to tissue dehydration), and stress at the incision was 0.97 ± 0.11.
CONCLUSIONS: No posterior capsule rupture was reported. Complete lens removal was achieved in 25 trials without significant surgical complications. Refinements to the procedures are required before fully automated lens extraction can be realized.
SETTING: Stein Eye Institute and Department of Mechanical and Aerospace Engineering, University of California, Los Angeles, USA.
DESIGN: Experimental study.
METHODS: Semiautomated lens extraction was performed on postmortem pig eyes using a robotic platform integrated with an OCT imaging system. Lens extraction was performed using a series of automated steps including robot-to-eye alignment, irrigation/aspiration (I/A) handpiece insertion, anatomic modeling, surgical path planning, and I/A handpiece navigation. Intraoperative surgical supervision and human intervention were enabled by real-time OCT image feedback to the surgeon via a graphical user interface. Manual preparation of the pig-eye models, including the corneal incision and capsulorhexis, was performed by a trained cataract surgeon before the semiautomated lens extraction procedures. A scoring system was used to assess surgical complications in a postoperative evaluation.
RESULTS: Complete lens extraction was achieved in 25 of 30 eyes. In the remaining 5 eyes, small lens pieces (≤1.0 mm3) were detected near the lens equator, where transpupillary OCT could not image. No posterior capsule rupture or corneal leakage occurred. The mean surgical duration was 277 seconds ± 42 (SD). Based on a 3-point scale (0 = no damage), damage to the iris was 0.33 ± 0.20, damage to the cornea was 1.47 ± 0.20 (due to tissue dehydration), and stress at the incision was 0.97 ± 0.11.
CONCLUSIONS: No posterior capsule rupture was reported. Complete lens removal was achieved in 25 trials without significant surgical complications. Refinements to the procedures are required before fully automated lens extraction can be realized.