MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2019.06.020

Медицинская библиотека · Journal of cataract and refractive surgery · 2019

Semiautomated optical coherence tomography-guided robotic surgery for porcine lens removal.

Chen Cheng-Wei, Francone Anibal Andrés, Gerber Matthew J, Lee Yu-Hsiu, Govetto Andrea, Tsao Tsu-Chin, Hubschman Jean-Pierre

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate semiautomated surgical lens extraction procedures using the optical coherence tomography (OCT)-integrated Intraocular Robotic Interventional Surgical System.

SETTING

Stein Eye Institute and Department of Mechanical and Aerospace Engineering, University of California, Los Angeles, USA.

DESIGN

Experimental study.

METHODS

Semiautomated lens extraction was performed on postmortem pig eyes using a robotic platform integrated with an OCT imaging system. Lens extraction was performed using a series of automated steps including robot-to-eye alignment, irrigation/aspiration (I/A) handpiece insertion, anatomic modeling, surgical path planning, and I/A handpiece navigation. Intraoperative surgical supervision and human intervention were enabled by real-time OCT image feedback to the surgeon via a graphical user interface. Manual preparation of the pig-eye models, including the corneal incision and capsulorhexis, was performed by a trained cataract surgeon before the semiautomated lens extraction procedures. A scoring system was used to assess surgical complications in a postoperative evaluation.

RESULTS

Complete lens extraction was achieved in 25 of 30 eyes. In the remaining 5 eyes, small lens pieces (≤1.0 mm3) were detected near the lens equator, where transpupillary OCT could not image. No posterior capsule rupture or corneal leakage occurred. The mean surgical duration was 277 seconds ± 42 (SD). Based on a 3-point scale (0 = no damage), damage to the iris was 0.33 ± 0.20, damage to the cornea was 1.47 ± 0.20 (due to tissue dehydration), and stress at the incision was 0.97 ± 0.11.

CONCLUSIONS

No posterior capsule rupture was reported. Complete lens removal was achieved in 25 trials without significant surgical complications. Refinements to the procedures are required before fully automated lens extraction can be realized.

Для цитирования: Chen Cheng-Wei, Francone Anibal Andrés, Gerber Matthew J, Lee Yu-Hsiu, Govetto Andrea, Tsao Tsu-Chin, Hubschman Jean-Pierre. Semiautomated optical coherence tomography-guided robotic surgery for porcine lens removal.. Journal of cataract and refractive surgery. 2019. DOI: 10.1016/j.jcrs.2019.06.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.