MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/bot.0b013e3182254ea3

Медицинская библиотека · Journal of orthopaedic trauma · 2012

A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures.

Forward Daren P, Doro Christopher J, OʼToole Robert V, Kim Hyunchul, Floyd John C P, Sciadini Marcus F, Turen Clifford H, Hsieh Adam H, Nascone Jason W

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To compare the biomechanical performance of a cephalomedullary nail (CMN), a proximal femoral locking plate, and a 95° angled blade plate in a comminuted subtrochanteric fracture model.

METHODS

A comminuted subtrochanteric femoral fracture model was created with a 2-cm gap below the lesser trochanter in 15 pairs of human cadaveric femora confirmed to be nonosteoporotic. The femora were randomized to treatment with one of the previously mentioned 3 devices. Each was tested under incrementally increasing cyclic load up to 90,000 cycles from 50% to 250% of body weight to simulate progressive weight bearing during 3 months of an average 700-N (approximately, 70 kg or 150 lb) person. Force, number of cycles, and total load sustained to reach 10 mm of displacement were compared. Failure modes were also noted.

RESULTS

The CMN construct withstood significantly more cycles, failed at a significantly higher force, and withstood a significantly greater load than either of the plate constructs (P < 0.001). Varus collapse was significantly lower in the CMN construct (P < 0.0001). Modes of failure differed among implant-bone constructs with damage to the femoral head through implant cutout in 5 of 10 blade plate specimens and 2 of 10 CMN specimens, whereas no damage to the femoral head bone was observed in any of the locking plate constructs.

CONCLUSIONS

The CMN construct was biomechanically superior to either the locking plate or 95° blade plate constructs. The locking plate construct was biomechanically equivalent to the blade plate construct.

Для цитирования: Forward Daren P, Doro Christopher J, OʼToole Robert V, Kim Hyunchul, Floyd John C P, Sciadini Marcus F, Turen Clifford H, Hsieh Adam H, Nascone Jason W. A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures.. Journal of orthopaedic trauma. 2012. DOI: 10.1097/bot.0b013e3182254ea3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.