Biomechanical Evaluation of Extramedullary Versus Intramedullary Reduction in Unstable Femoral Trochanteric Fractures.

Kawamura Tadashi, Minehara Hiroaki, Tazawa Ryo, Matsuura Terumasa, Sakai Rina, Takaso Masashi

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

The failure rate of operations involving the cephalomedullary nail technique for unstable femoral trochanteric fractures is 3-12%. Changing the reduction strategy may improve the stability. This study aimed to confirm whether reducing the proximal fragment with the medial calcar contact, as opposed to utilizing an intramedullary reduction, would improve the stability of such fractures.

MATERIALS AND

METHODS

The unstable femoral trochanteric fracture model was created with fixation by cephalomedullary nails in 22 imitation bones. The 2 reduction patterns were as follows: one was with the proximal head-neck fragment external to the distal bone in the frontal plane and anterior in the sagittal plane as "Extramedullary," while the other was the opposite reduction position, that is, bone in the frontal plane and sagittal plane as "Intramedullary." We evaluated the tip-apex distance, compression stiffness, change in femoral neck-shaft angle, amount of blade telescoping, and diameter of the distal screw hole after the compression test. Statistical analysis was conducted using the Mann-Whitney U test.

RESULTS

No significant differences were seen in compression stiffness (p = 0.804) and femoral neck-shaft angle change (p = 0.644). Although the "Extramedullary" tip-apex distance was larger than the "Intramedullary" distance (p = 0.001), it indicated clinically acceptable lengths. The amount of blade telescoping and the distal screw hole diameter were significantly larger in "Intramedullary" than in "Extramedullary" (p < 0.001, p = 0.019, respectively). Our results showed that "Intramedullary" had significantly larger blade telescoping and distal screw hole diameters than "Extramedullary," and contrary to our hypothesis, no significant differences were seen in compression stiffness and femoral neck-shaft angle change.

CONCLUSIONS

As opposed to the "Intramedullary" reduction pattern, the biomechanical properties of the "Extramedullary" reduction pattern improved stability during testing and decreased sliding.

Для цитирования: Kawamura Tadashi, Minehara Hiroaki, Tazawa Ryo, Matsuura Terumasa, Sakai Rina, Takaso Masashi. Biomechanical Evaluation of Extramedullary Versus Intramedullary Reduction in Unstable Femoral Trochanteric Fractures.. Geriatric orthopaedic surgery & rehabilitation. 2021. DOI: 10.1177/2151459321998611.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.