Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry.

Xu Andrew, Francis Bincy, Choy Kay Weng, Dargaville George Francis, Surkitt Amy, Tran David, Subhi Rami, Fan Wei Qi

СтатьяПеревод в очереди

Аннотация

Background

Neonatal jaundice is a common condition with potentially severe complications such as bilirubin-induced neurological dysfunction and kernicterus. While serum bilirubin (SBR) remains the standard laboratory measurement, point-of-care methods, such as transcutaneous bilirubinometry (TcB) and blood gas analysers (BGAs), offer rapid, less invasive alternatives. Direct comparisons of their diagnostic accuracy remain limited.

Objective

The aim of this study was to assess and compare diagnostic accuracy and clinical utility of TcB and BGA against SBR in neonatal hyperbilirubinaemia screening.

Methods

This retrospective study included neonates (n = 221) with concurrent SBR, BGA, and TcB measurements (n = 333). Assessment was via Passing-Bablok regression, Bland-Altman analysis, and Spearman correlation. Diagnostic performance was evaluated against jaundice thresholds in phototherapy charts (≥95th percentile threshold). Subgroup analyses considered phototherapy status, haemoglobin concentration, and Fitzpatrick skin type.

Results

BGA showed stronger agreement with SBR (R2 = 0.88) than TcB (R2 = 0.43). BGA remained accurate regardless of phototherapy or haemoglobin levels. TcB accuracy declined post-phototherapy with reduced predictive value in darker-skinned neonates (Fitzpatrick III-VI) and increased false discovery rates. Both methods demonstrated low sensitivity (45.8%) but high specificity (>95%) and negative predictive value (~91%) for clinically significant hyperbilirubinaemia. BGA had a higher diagnostic odds ratio (47.5) than TcB (19.3). When individual patient sequential SBR and BGA measurements were compared for jaundice tracking (n = 175), there was high correlation, (r = 0.971) with no statistical differences, and 50% of measurements achieving agreement within 10 μmol/L.

Conclusions

BGA is a more reliable alternative to SBR than TcB, particularly in time-critical or resource-limited settings. While TcB remains a non-invasive screening tool, limited accuracy post-phototherapy and with darker skinned neonates indicate confirmatory SBR testing. These findings support the selective and context-aware use of BGA and TcB to optimise neonatal hyperbilirubinaemia management and reduce interventions.

Для цитирования: Xu Andrew, Francis Bincy, Choy Kay Weng, Dargaville George Francis, Surkitt Amy, Tran David, Subhi Rami, Fan Wei Qi. Point-of-Care Bilirubin Testing in Neonates: Comparative Performance of Blood Gas Analysis and Transcutaneous Bilirubinometry.. Healthcare (Basel, Switzerland). 2026. DOI: 10.3390/healthcare14030370.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.