The mechanical properties of breast prostheses.

Peters W J

СтатьяПеревод в очереди

Аннотация

The mechanical properties of inflatable and gel-filled breast prostheses were evaluated using the Instron Universal Testing Machine. Prosthesis strength characteristics were evaluated in terms of compression strength (rather than tensile strength) because of the relationship to closed capsulotomy. The compression breaking strength of prostheses ranged from 0.62 to 10.8 pounds per square inch. There was considerable variation among prostheses. Pressures exceeding these values have been recorded during closed compression capsulotomy. The clinical relevance of these results is discussed.