Biocompatibility of a new device of self-expandable covered and non-covered tracheal stent: comparative study in rats.

Rodrigues Olavo Ribeiro, Minamoto Hélio, Canzian Mauro, Correia Aristides Tadeu, Jatene Fabio Biscegli

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To investigate the compatibility of a new model of self-expandable tracheal stent in rats.

METHODS

A new device of polyurethane covered and non - covered stent was placed in the trachea of Wistar rats. Animals were distributed in two groups: the polyurethane covered and non-covered group. Macroscopic parameters included position within the tracheal lumen, adherence to the mucosa, degree of dilatation, permeability and internal diameter. Microscopic findings evaluated were: incorporation, inflammatory activity, granulation tissue and epithelial revetment injuries. The observation follow-up was six weeks. All parameters were quantified based on determined score values. Incorporation of the stents was evaluated based on the observation if the stent was fixed into the trachea or if it could be removed. Degree of dilatation was performed by external diameter measurements. Granulation tissue was evaluated by measurements of height of the tissue growing into the tracheal lumen.

RESULTS

100% of non-covered stents had total attachment to mucosa and 100% of polyurethane covered type had adherence only. Regarding dilatation, granulation tissue, inflammatory activity and internal diameter measurements, there were no significant differences between the groups. Pathological tracheal wall injuries were present in both groups.

CONCLUSION

Both models of stent demonstrated biocompatibility with the trachea. Rats are suitable for an experimental model of tracheal stent study.

Для цитирования: Rodrigues Olavo Ribeiro, Minamoto Hélio, Canzian Mauro, Correia Aristides Tadeu, Jatene Fabio Biscegli. Biocompatibility of a new device of self-expandable covered and non-covered tracheal stent: comparative study in rats.. Acta cirurgica brasileira. 2013. DOI: 10.1590/s0102-86502013000100003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.