A novel in-line high frequency interrupter for use with bubble CPAP: A feasibility study in a premature lamb model.

Sivieri E M, Eichenwald E C, Abbasi S, Wolfson M R

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Recent in vitro testing of high frequency (HF) oscillation applied to bubble continuous positive airway pressure (BCPAP) using a novel flow interrupter device (HFI) demonstrated significantly improved CO2 washout while not altering delivered mean airway pressure (MAP) in a premature infant lung model. This study's aim was to evaluate the safety and efficacy of the HFI paired with BCPAP in an animal model of prematurity prior to clinical testing.

DESIGN/

METHODS

Twelve fetal lambs, 131-135 days gestation, weight 3.51±0.42 kg, were delivered by Cesarean section. The lambs were supported by mechanical ventilation and weaned to spontaneous breathing with BCPAP at 6 cmH2O. A combined CO2/airflow sensor measured end-tidal (EtCO2) and tidal volume (VT). Blood gases, heart rate (HR), arterial pressure (Part), minute ventilation (MV), MAP, ventilatory efficiency index (VEI), thoracoabdominal phase angle and labored breathing index (LBI) were recorded over a 10-minute baseline period followed by four randomized 10-minute intervals with HFI set to either 8, 10, 12 or 15 Hz.

RESULTS

EtCO2 decreased from baseline by 11.1±2.2SE%, 16.6±4.3SE%, 13.5±4.9SE%, and 19.5±4.5SE% at 8, 10, 12, and 15 Hz respectively (p < 0.001). Blood gases, SpO2, HR, Part, MAP, VT, MV, esophageal pressure, phase angle, and LBI underwent no significant change with HF. Respiratory rate decreased, and VEI increased, by 14.9±4.5SD% (p = 0.037) and 83±22SD% (p < 0.011) respectively, averaged over all frequencies.

CONCLUSIONS

We demonstrated the safety and efficacy of a novel BCPAP flow interrupter device. HF applied to the respiratory system resulted in significantly improved CO2 clearance and ventilation efficiency with no deleterious physiological effects in a pre-term lamb model.

Для цитирования: Sivieri E M, Eichenwald E C, Abbasi S, Wolfson M R. A novel in-line high frequency interrupter for use with bubble CPAP: A feasibility study in a premature lamb model.. Journal of neonatal-perinatal medicine. 2022. DOI: 10.3233/npm-210899.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.