DOI 10.3760/cma.j.cn121094-20221026-00514
[Determination of Perchloroethylene in blood by headspace gas chromatography-mass spectrometry].
Аннотация
Objective
To establish a method for determination of Perchloroethylene (PCE) in blood by headspace gas chromatography-mass spectrometry (HS/GC-MS) .
Methods
From Dctober to December 2021, A total of 3 mL blood samples were taken into a 10 mL headspace bottle, after heated at 60 ℃ for 30 mins, PCE in the top air was separated by VF-WAXms capillary column and detected by GC-MS. The retention time and external standard method were used for qualitative and quantitative analysis of PCE in samples, respectively.
Results
There was good linear relationship in the range of 5.09-200.17 μg/L. The linear correlation coefficient was 0.9993.The detection limit was 0.21 μg/L and the lower limit of quantitation was 0.70 μg/L. The recovery rates of samples with different concentrations were 95.3%-103.8%. The intra-batch relative standard deviations (RSD) were 3.2%-4.6%, and inter-batch RSD was 4.0%-6.1%. The samples can be stored at 4 ℃ for three days and at -20 ℃ for seven days.
Conclusion
This method is proved to be simple, practical and highly sensitive, which is suitable for the determination of PCE in blood.