Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2018
Safety and efficacy of the new bidirectional rotational Evolution® mechanical lead extraction sheath: results from a multicentre Italian registry.
СтатьяМногоцентровое исследованиеНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mazzone Patrizio, Migliore Federico, Bertaglia Emanuele, Facchin Domenico, Daleffe Elisabetta, Calzolari Vittorio, Crosato Martino, Melillo Francesco, Peruzza Francesco, Marzi Alessandra, Sora Nicoleta, Della Bella Paolo
Аннотация
AIMS: The aim of this prospective multicentre study is to evaluate safety and efficacy of the new bidirectional rotational mechanical lead extraction (LE) sheath (Evolution RL, Cook Medical, USA) in chronically implanted leads (>1-year-old leads).
METHODS AND RESULTS: Between September 2013 and June 2016, a total of 238 leads in 124 consecutive patients were removed by using the new Evolution RL rotational mechanical sheath. Indications for LE were cardiac device infection in 63 (50.8%) cases, lead malfunction in 41 (33.1%), upgrade in 1 (0.8%) case and for other reasons in the remaining 19 cases (15.3%). Ninety-one leads (38.2%) were implantable cardioverter defibrillator leads (81 dual coil vs. 10 single coil), 38 (16%) right ventricular leads, 86 (36.1%) right atrial leads, and 23 (9.7%) coronary sinus leads. The mean implant duration was 92.2 ± 52.9 months (range 12-336). 91.6% of the leads (218/238) were extracted completely with the Evolution RL alone, with the complete success rate rising to 98.7% (235/238 leads) with combined use of a snare. Overall clinical success rate was 100%. No Evolution sheath-related complications were noted. There were no deaths or major complications. Five minor complications (4%) were encountered. In cases of companion leads no wrapping or lead damage were observed.
CONCLUSION: On the basis of our prospective multicentre study, the new hand-powered bidirectional rotational mechanical LE sheath is an effective and safe tool for the extraction of chronically implanted leads without major complications and lead wrapping or lead damage.
METHODS AND RESULTS: Between September 2013 and June 2016, a total of 238 leads in 124 consecutive patients were removed by using the new Evolution RL rotational mechanical sheath. Indications for LE were cardiac device infection in 63 (50.8%) cases, lead malfunction in 41 (33.1%), upgrade in 1 (0.8%) case and for other reasons in the remaining 19 cases (15.3%). Ninety-one leads (38.2%) were implantable cardioverter defibrillator leads (81 dual coil vs. 10 single coil), 38 (16%) right ventricular leads, 86 (36.1%) right atrial leads, and 23 (9.7%) coronary sinus leads. The mean implant duration was 92.2 ± 52.9 months (range 12-336). 91.6% of the leads (218/238) were extracted completely with the Evolution RL alone, with the complete success rate rising to 98.7% (235/238 leads) with combined use of a snare. Overall clinical success rate was 100%. No Evolution sheath-related complications were noted. There were no deaths or major complications. Five minor complications (4%) were encountered. In cases of companion leads no wrapping or lead damage were observed.
CONCLUSION: On the basis of our prospective multicentre study, the new hand-powered bidirectional rotational mechanical LE sheath is an effective and safe tool for the extraction of chronically implanted leads without major complications and lead wrapping or lead damage.