DOI 10.1016/j.ab.2021.114435

Медицинская библиотека · Analytical biochemistry · 2021

High-throughput and trace analysis of diazepam in plasma using DART-MS/MS and its pharmacokinetic application.

Liu Yingze, Su Chong, Zhang Yuyao, Zhang Di, Li Yaoshuang, Gu Jingkai, Wang Ensi, Sun Dong

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

A high-throughput quantitative analytical method based on Direct Analysis in Real Time tandem mass spectrometry (DART-MS/MS) has been developed and validated for the determination of diazepam in rat plasma, whereby analyzing of each sample needs merely 25 μL plasma, simple solid phase extraction sample preparation and 15 s acquisition time. The multiple reaction monitoring (MRM) transitions at m/z 285.2 → 193.1 and 316.0 → 270.0 were selected for the monitoring of diazepam and its internal standard clonazepam respectively. A good linearity within the range of 10-2000 ng/mL, an intra- and inter-day precisions within <7.78% as to an accuracy ranging from 1.04% to 7.92% have been achieved. The method has been successfully applied to the pharmacokinetic study of diazepam in rats' plasma after a single intragastric administration at a dose of 10 mg/kg. The results indicate that this method fulfills the requirements of the bioanalysis in sensitivity and accuracy. It shows considerable promise for application of DART-MS to the quantitative investigation of other drugs.

Для цитирования: Liu Yingze, Su Chong, Zhang Yuyao, Zhang Di, Li Yaoshuang, Gu Jingkai, Wang Ensi, Sun Dong. High-throughput and trace analysis of diazepam in plasma using DART-MS/MS and its pharmacokinetic application.. Analytical biochemistry. 2021. DOI: 10.1016/j.ab.2021.114435.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.